Key result
High-bandwidth force plate BCG improves pre-ejection period estimation and tracks stroke volume changes with ~5% error.
Why the study?
Does a high bandwidth force plate ballistocardiogram improve the accuracy of pre-ejection period and stroke volume change estimations compared to a modified weighing scale in healthy subjects?
Observational (n=17)
No
Does a high bandwidth force plate ballistocardiogram improve the accuracy of pre-ejection period and stroke volume change estimations compared to a modified weighing scale in healthy subjects?
A high-bandwidth force plate BCG provides superior timing accuracy for non-invasively estimating pre-ejection period and relative stroke volume changes compared to a modified weighing scale, establishing a framework for unobtrusive out-of-clinic hemodynamic monitoring.
Answers are grounded in this paper and cite their sources.
Supports force-plate BCG development for out-of-clinic monitoring; leaves open validation in clinical populations.
Unobtrusive and inexpensive technologies for monitoring the cardiovascular health of heart failure (HF) patients outside the clinic can potentially improve their continuity of care by enabling therapies to be adjusted dynamically based on the changing needs of the patients. Specifically, cardiac contractility and stroke volume (SV) are two key aspects of cardiovascular health that change significantly for HF patients as their condition worsens, yet these parameters are typically measured only in hospital/clinical settings, or with implantable sensors. In this work, we demonstrate accurate measurement of cardiac contractility (based on pre-ejection period, PEP, timings) and SV changes in subjects using ballistocardiogram (BCG) signals detected via a high bandwidth force plate. The measurement is unobtrusive, as it simply requires the subject to stand still on the force plate while holding electrodes in the hands for simultaneous electrocardiogram (ECG) detection. Specifically, we aimed to assess whether the high bandwidth force plate can provide accuracy beyond what is achieved using modified weighing scales we have developed in prior studies, based on timing intervals, as well as signal-to-noise ratio (SNR) estimates. Our results indicate that the force plate BCG measurement provides more accurate timing information and allows for better estimation of PEP than the scale BCG (r² = 0.85 vs. r² = 0.81) during resting conditions. This correlation is stronger during recovery after exercise due to more significant changes in PEP (r² = 0.92). The improvement in accuracy can be attributed to the wider bandwidth of the force plate. ∆SV (i.e., changes in stroke volume) estimations from the force plate BCG resulted in an average error percentage of 5.3% with a standard deviation of ±4.2% across all subjects. Finally, SNR calculations showed slightly better SNR in the force plate measurements among all subjects but the small difference confirmed that SNR is limited by motion artifacts rather than instrumentation.
No takes yet. Share an insight, caveat, or question.
Ashouri et al. (2016) conducted an observational in Healthy subjects (technology intended for heart failure monitoring) (n=17). Ballistocardiogram (BCG) measurement via a high bandwidth multi-component force plate (Kistler Type 9260AA6) vs. Modified electronic weighing scale (Tanita BC534) BCG and reference impedance cardiography (ICG) was evaluated on Correlation of BCG RI-interval with reference pre-ejection period (PEP) from impedance cardiography during rest. High-bandwidth force plate BCG improved pre-ejection period estimation over scale BCG at rest and post-exercise, estimating stroke volume changes with 5.3% mean error.