Key result
Automatic correction algorithm cuts catheter pressure tracing error by ~5.1 mmHg vs. reference wire.
Why the study?
Errors, corrections, and clinical implications of fluid-filled catheter phasic pressure assessments have not been assessed in the era of high-fidelity pressure wire sensors.
Does a 5-parameter automatic software correction algorithm (accounting for offset, timing, gain, and harmonic oscillations) reduce the root mean square error between fluid-filled guide catheter and high-fidelity 0.014-inch pressure wire phasic aortic pressure tracings in patients undergoing invasive coronary or valvular physiology assessment?
Cohort (n=23)
No
Does a 5-parameter automatic software correction algorithm (accounting for offset, timing, gain, and harmonic oscillations) reduce the root mean square error between fluid-filled guide catheter and high-fidelity 0.014-inch pressure wire phasic aortic pressure tracings in patients undergoing invasive coronary or valvular physiology assessment?
Absolute Event Rate: 0.52% vs 5.6%
p-value: p=<0.001
An automated 5-parameter calibration algorithm corrects fluid-filled guide catheter pressure artifacts (offset, timing, gain, and harmonic ringing) to within <1 mmHg RMS error of a high-fidelity pressure wire, improving the precision of subcycle coronary and valvular hemodynamic indices without additional hardware.
May refine fluid-filled catheter precision in physiology labs; leaves open whether it affects decisions or outcomes in prospective use.
OBJECTIVES: We sought to develop an automatic method for correcting common errors in phasic pressure tracings for physiology-guided interventions on coronary and valvular stenosis. BACKGROUND: Effective coronary and valvular interventions rely on accurate hemodynamic assessment. Phasic (subcycle) indexes remain intrinsic to valvular stenosis and are emerging for coronary stenosis. Errors, corrections, and clinical implications of fluid-filled catheter phasic pressure assessments have not been assessed in the current era of ubiquitous, high-fidelity pressure wire sensors. METHODS: We recruited patients undergoing invasive coronary physiology assessment. Phasic aortic pressure signals were recorded simultaneously using a fluid-filled guide catheter and 0.014″ pressure wire before and after standard calibration as well as after pullback. We included additional subjects undergoing hemodynamic assessment before and after transcatheter aortic valve implantation. Using the pressure wire as reference standard, we developed an automatic algorithm to match phasic pressures. RESULTS: Removing pressure offset and temporal shift produced the largest improvements in root mean square (RMS) error between catheter and pressure wire signals. However, further optimization <1 mmHg RMS error was possible by accounting for differential gain and the oscillatory behavior of the fluid-filled guide. The impact of correction was larger for subcycle (like systole or diastole) versus whole-cycle metrics, indicating a key role for valvular stenosis and emerging coronary pressure ratios. CONCLUSIONS: When calibrating phasic aortic pressure signals using a pressure wire, correction requires these parameters: offset, timing, gain, and oscillations (frequency and damping factor). Automatically eliminating common errors may improve some clinical decisions regarding physiology-based intervention.
No takes yet. Share an insight, caveat, or question.
Johnson et al. (2020) conducted a cohort in Coronary artery stenosis (with a supplementary cohort of aortic stenosis) (n=23). 5-parameter automatic phasic pressure correction algorithm (adjusting for offset, timing, differential gain, and harmonic oscillator frequency and damping) vs. Raw fluid-filled guide catheter tracing and standard pressure offset equalization using a 0.014-inch high-fidelity pressure wire as reference was evaluated on Root mean square (RMS) error (mmHg) between fluid-filled guide catheter and 0.014-inch reference pressure wire aortic pressure signals (p=<0.001). The 5-parameter automatic correction algorithm reduced median root mean square error between fluid-filled catheter and pressure wire tracings from 5.6 mmHg to 0.52 mmHg (p<0.001).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: