Answers are grounded in this paper and cite their sources.
A new device for local delivery of antibiotics is presented, with potential use as a drug-eluting fixation pin for orthopedic applications. The implant consists of a stainless steel hollow tubular reservoir packed with the desired antibiotic. Release takes place through several orifices previously drilled in the reservoir wall, a process that does not compromise the mechanical properties required for the implant. Depending on the antibiotic chosen and the number of orifices, the release profile can be tailored from a rapid release of the load (ca. 20h) to a combination of rapid initial release and slower, sustained release for a longer period of time (ca. 200h). An excellent bactericidal action is obtained, with 4-log reductions achieved in as little as 2h, and total bacterial eradication in 8h using 6-pinholed implants filled with cefazolin.
No takes yet. Share an insight, caveat, or question.
Gimeno et al. (2015) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: