The theory and design considerations for dynamic condenser electrometers are presented. The main features of such instruments are simple and dependable operation, extremely low drift and high sensitivity. (10,000 divisions/volt is easily obtainable.) Also included is the description and performance of two instruments specifically designed to measure d.c. currents of the order of 10−15 ampere.