Answers are grounded in this paper and cite their sources.
A peroxidase model electrode was devised for reagentless sensing of hydrogen peroxide (H2O2). A small model molecule, which mimics the vicinity of the reaction center of a redox enzyme, can communicate electrochemically with an electrode. Heme nonapeptide (MW congruent to 1600) having peroxidase activity was adopted as a peroxidase model compound and was covalently immobilized on a tin oxide (SnO2) electrode as a roughly monomolecular layer. The modified electrode thus obtained responded to H2O2 at concentrations down to 10(-6) M without electron mediator or promoter, at a mild potential of +150 or +300 mV vs Ag/AgCl. In a batch system, the response reached a steady state in a few seconds. Measurements were possible also in a flow system with an assay time of 0.5-1.0 min/sample. The steady-state response of the electrode was kinetically analyzed.
No takes yet. Share an insight, caveat, or question.
Tatsuma et al. (1991) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: