Experimental study demonstrates color correction using color matching functions achieves over 95% accuracy in portable urine analysis, indicating reliable device-independent point-of-care testing.
Color correction methods of chromaticity coordinates using Color Matching Function (CMF) were studied to develop a device-independent portable-type urine analyzer. The reflection spectra were measured for the degrees of 10 test items of the urine reagent strip (urine strip) to develop a portable-type urine analyzer. A computer simulation was performed to quantitatively distinguish the color reactions of the urine system, by using the spectral power distribution of Light Emitting Diode(LED), the reflection of a urine strip, and spectral sensitivity of a photodiode. To develop a device-independent system, chromaticity coordinates were modified to reduce the color deviations in the urine strip, by using the temperature compensation of LED and the color transformation by CMF. The experimental values obtained by developed urine system exhibited the accuracy above 95% for all color samples.
No takes yet. Share an insight, caveat, or question.
Kim et al. (2002) studied this question.