This is the authors' abstract. We don't add key points for this paper.
The characteristics of the cylindrical mirror analyzer (CMA) attached to a commercial-type scanning Auger electron microprobe, JAMP-3, were examined for its possible use in monitoring the variation of surface properties through the measurement of the energy distribution of secondary electrons, the N s ( E )-spectrum. For this purpose, a special compact electron gun, small enough to be attached inside a sample holder, was made to calibrate the energy scale of the CMA, and a special arrangement for sample setting was devised to measure the N s ( E )-spectra. This arrangement enables the variation of surface potential to be detected throught the shift of the cut-off point of the N s ( E )-spectrum with as good accuracy as can be obtained by an ESCA-system.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Ogoh et al. (1985) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: