Answers are grounded in this paper and cite their sources.
We demonstrate intravascular optical coherence tomography (OCT) imaging with frame rate up to 3.2 kHz (192,000 rpm scanning). This was achieved by using a custom-built catheter in which the circumferential scanning was actuated by a 1.0 mm diameter synchronous motor. The OCT system, with an imaging depth of 3.7 mm (in air), is based on a Fourier domain mode locked laser operating at an A-line rate of 1.6 MHz. The diameter of the catheter is 1.1 mm at the tip. Ex vivo images of human coronary artery (78.4 mm length) were acquired at a pullback speed of 100 mm/s. True 3D volumetric imaging of the entire artery, with dense and isotropic sampling in all dimensions, was performed in <1 second acquisition time.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2013) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: