Key result
Manumeter HAND algorithm tracks home hand movement with ~85% accuracy.
Observational (n=29)
Answers are grounded in this paper and cite their sources.
There are few wearable sensors suitable for daily monitoring of wrist and finger movements for hand-related healthcare applications. Here, we describe the development and validation of a novel algorithm for magnetically counting hand movements. We implemented the algorithm on a wristband that senses magnetic field changes produced by movement of a magnetic ring worn on the finger (the "Manumeter"). The "HAND" (Hand Activity estimated by Nonlinear Detection) algorithm assigns a "HAND count" by thresholding the real-time change in magnetic field created by wrist and/or finger movement. We optimized thresholds to achieve a HAND count accuracy of ~85% without requiring subject-specific calibration. Then, we validated the algorithm in a dexterity-impaired population by showing that HAND counts strongly correlate with clinical assessments of upper extremity (UE) function after stroke. Finally, we used HAND counts to test a recent hypothesis in stroke rehabilitation that real-world UE hand use increases only for stroke survivors who achieve a threshold level of UE functional capability. For 29 stroke survivors, HAND counts measured at home did not increase until the participants' Box and Blocks Test scores exceeded ~50% normal. These results show that a threshold-based magnetometry approach can non-obtrusively quantify hand movements without calibration and also verify a key concept of real-world hand use after stroke.
No takes yet. Share an insight, caveat, or question.
Lucena et al. (2021) conducted an observational in Stroke with upper extremity dexterity impairment (n=29). Manumeter with calibration-free HAND algorithm vs. Clinical assessments of upper extremity function (Box and Blocks Test) was evaluated on HAND count accuracy and real-world hand use correlation with Box and Blocks Test scores. The Manumeter HAND algorithm achieved ~85% hand movement counting accuracy and showed home hand use increased only when Box and Blocks Test scores exceeded ~50% normal.
Synapse has enriched 10 related papers. Consider them for comparative context: