Nondispersive infrared (NDIR) CO 2 analyzer could produce erroneous CO 2 mole fraction measurements for an air sample when CO 2 ‐in‐air mixtures having different isotopic compositions than atmospheric CO 2 are used as the NDIR calibration gases. This is because (1) an optical band‐pass filter equipped in a typical NDIR analyzer to minimize the interference effect from the other infrared‐active species is basically designed to transmit only the absorption band of 12 C 16 O 2 and (2) absorption bands for the other CO 2 ‐related isotopologues, for example, 13 C 16 O 2 , are shifted to lower wave numbers depending on their isotope effects. To evaluate the effect of the isotopic composition on the NDIR response, we computed the theoretical relative molar response of the instrument to each isotopologue based on the infrared absorptance by the individual isotopologues. We then prepared a gravimetric 13 CO 2 ‐in‐air mixture with CO 2 mole fraction of 380 ppm to experimentally determine the optical filter property. The apparent mole fractions of the 13 CO 2 ‐in‐air mixture determined by three NDIR analyzers used in this study were 46, 94, and 27 ppm, indicating that the optical filters in these instruments substantially reduced the response to 13 C 16 O 2 . Based on these theoretical and experimental analyses, we evaluated the apparent difference in the CO 2 mole fraction determined by the three NDIR analyzers from the true value when isotopically lighter CO 2 ‐in‐air mixtures ( δ 13 C = −32.4‰ and δ 18 O = +11.7‰), as compared to atmospheric CO 2 ( δ 13 C = ∼−8‰ and δ 18 O = ∼+40‰), are used as calibration gases. The estimated difference varied with NDIR analyzers, ranging from −0.04 to −0.08 ppm for air samples with 380 ppm CO 2 .
No takes yet. Share an insight, caveat, or question.
Tohjima et al. (2009) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: