Answers are grounded in this paper and cite their sources.
Background: In spinal fusion surgery, intersomatic compression force is currently applied subjectively by the operating surgeon, despite its critical role on implant stability and risk of subsidence. No standardized measurement or guideline exists to control or quantify the amount of force applied. Methods: In a two-phase exploratory study, we evaluated whether proprioceptive muscle memory allows reliable reproduction of applied manual compression forces. In Phase 1, 30 participants applied force to a compression clamp equipped with a strain gauge, simulating spinal interbody compression on a 3D-printed vertebral model. They were then asked to reproduce this force using a hand dynamometer at defined time intervals. In Phase 2, intraoperative compression forces applied during spinal fusion procedures were retrospectively assessed by having the operating surgeon reproduce the force on a dynamometer. Results: Participants were able to reproduce their initial manual compression force within a 15% deviation, even 15 min after the initial application. In 116 clinical cases, an average compression force of 146.3 ± 18.5 N was recorded. No significant differences were observed across different spinal segments. Conclusions: These findings provide initial data toward defining a reproducible reference range for indirect intraoperative compression assessment. Standardization of applied force may help improve biomechanical outcomes and reduce complications such as implant migration, pseudarthrosis, or cage subsidence.
No takes yet. Share an insight, caveat, or question.
Heilmann et al. (2025) studied this question.