Answers are grounded in this paper and cite their sources.
Summary Combination GC-MS interfacial systems have now been developed to a stage where it is possible to transfer the components of complex mixtures unchanged and without loss in GC resolution. It would, however, appear that no one separator is completely satisfactory for all applications, particularly if the solutes under examination are of low molecular weight and present in submicrogram quantities. In the author's opinion the following three interfaces would satisfy all possible requirements: 1. Palladium separator – specifically for low-molecular-weight volatile inorganic compounds and compounds that would not be reduced by the presence of active hydrogen. 2. Micro silicone rubber membrane separator for organic compounds up to a molecular weight of 250. 3. Jet separator – single-stage (variable) or double-stage, dependent on carrier-gas flow rate, for molecular weights from 100 upwards.
No takes yet. Share an insight, caveat, or question.
Simpson et al. (1972) studied this question.