Answers are grounded in this paper and cite their sources.
Nitrous acid (HONO) plays an important role in the formation of hydroxyl radical and the nitrogen cycle in atmospheric chemistry. Cavity ring-down spectroscopy (CRDS) has been utilized for detection of gaseous HONO in the near-UV spectral region. At the maximum absorption wavelength (354.2 nm), HONO can be detected with a low limit of 5 parts per billion (ppb) in 15 s sampling time, while at a low absorption wavelength, with a high value of 10 parts per million (ppm). The CRDS technique has a large dynamic range of detection, with a good linearity of absorbance versus concentration. The detection sensitivity demonstrated in this study is in the range of the ambient HONO concentration (0.1−10 ppb), and it can be readily improved to a sub-ppb level with upgraded optics and longer cavity. Moreover, this high detection sensitivity is realized with a relatively compact setup (in ∼ meters) and with a short sampling time (in 15 s). CRDS renders a promising technique for real-time and absolute measurements of HONO in laboratory and ambient studies. Absorption spectrum and cross sections of HONO near 354 nm have also been examined.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2000) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: