Key result
Intra-arterial coil stenting preserves a ~0.4-mm larger luminal diameter at 4 weeks versus dilatation alone.
Answers are grounded in this paper and cite their sources.
Demonstrates that stent placement reduces restenosis and preserves luminal diameter in atherosclerotic rabbits, suggesting improved vascular outcomes over balloon angioplasty alone.
The major problem associated with percutaneous transluminal coronary angioplasty is recurrence of the stenotic lesion. Balloon catheter-mounted intracoronary stent devices may reduce restenosis by improving luminal morphology and flow characteristics. This study assessed the effects of a stainless steel wire, interdigitating coil stent on restenosis in the atherosclerotic rabbit model. Fifteen cholesterol-fed rabbits with preexisting iliac arterial lesions induced by balloon deendothelialization were instrumented in one iliac artery with a 2.0-mm diameter stent after balloon dilatation; the contralateral iliac lesion was treated by dilatation only to serve as a control. The animals were given heparin with aspirin (60 mg) and aspirin every 3rd day until death. Arteriography was repeated 4 weeks after stenting, just before death. Tissue sections from stented and control arterial segments were analyzed morphometrically. Stented arteries had a significantly larger luminal diameter at restudy, whether measured by arteriography (1.38 +/- 0.19 vs. 0.94 +/- 0.35 mm, p less than 0.01) or from tissue sections (1.26 +/- 0.18 vs. 0.81 +/- 0.30 mm, p = 0.0001). Wall thickness of the stented segment was slightly, but significantly, less than the control segment (436 +/- 143 vs. 532 +/- 221 microns, p less than 0.05). Scanning electron microscopy of five stented atherosclerotic rabbit aortas revealed regeneration of a nonthrombogenic, confluent, flow-directed endothelium by 4 weeks after placement. Intra-arterial stenting may be of benefit in the prevention of restenosis by the preservation of a larger functional lumen and by a decrease in the neointimal hyperplastic response to arterial injury.
No takes yet. Share an insight, caveat, or question.
Robinson et al. (1988) studied Atherosclerotic iliac arterial lesions (restenosis animal model) (n=15). Stainless steel wire interdigitating coil stent (2.0-mm diameter) after balloon dilatation vs. Contralateral iliac lesion treated by balloon dilatation only was evaluated on Luminal diameter at restudy (arteriography, mm) (p=<0.01). Intra-arterial coil stenting preserved a larger luminal diameter at 4 weeks vs dilatation alone by arteriography and morphometry.
Synapse has enriched 10 related papers. Consider them for comparative context: