A double-sided, high-frequency response heat-flux gauge has been developed which allows measurement of heat flux from dc to 100 kHz. The instrument is designed for heat-flux magnitudes ranging from one to several hundred kW/m2 at temperatures up to 400 °C and is independent of the test article material. The gauges consist of a metal film (1500 Å) resistance thermometers sputtered on both sides of a thin (25 μm) polyimide sheet. The sheet, which can contain many gauges, is then adhesively bonded to a test article. The temperature difference across the polyimide is a direct measure of the heat flux at low frequencies, while a quasi-1D analysis is used to infer the high-frequency heat flux from the upper surface temperature history. The design criteria, construction and application techniques, and a novel, ratiometric calibration procedure are discussed in detail.