A design of a spectrometer system with a fixed self-focusing grating and a linear diode array (Reticon 512G) is described, where a microprocessor directly evaluates the pressure in realtime from ruby luminescence spectra at a typical repetition rate of 2 scans per second and a precision of better than +or-0.025 GPa over a range of nearly 50 GPa. A self-calibration procedure allows for reliability tests of the instrument at any time. As an application a time-resolved isochoric study of the pressure behaviour at the phase boundary between ice VI and the liquid is shown.
No takes yet. Share an insight, caveat, or question.
Hirsch et al. (1983) studied this question.