Key result
Active hemoperfusion during PTCA can cause coronary pseudo-occlusion that resolves immediately upon termination.
Case Report
Answers are grounded in this paper and cite their sources.
We describe the phenomenon of coronary artery "pseudo-occlusion," which may occur when active hemoperfusion devices are utilized during percutaneous transluminal coronary angioplasty (PTCA). In such cases, contrast injected via the guiding catheter fails to opacify the coronary artery distal to the tip of the dilating catheter, thereby giving the angiographic appearance of coronary occlusion. When active hemoperfusion is terminated, contrast opacifies the vessel in a normal fashion. The cause of this observation is likely multifactorial and includes obstruction to flow by the dilating catheter, high distal pressure delivered by the dilating catheter during hemoperfusion competing with the flow of contrast injected via the guiding catheter, and wash-out of contrast at the tip of the dilating catheter by non-contrast containing blood being pumped through the system. This phenomenon must be recognized if potentially deleterious repeat dilatations are to be avoided.
No takes yet. Share an insight, caveat, or question.
Grill et al. (1991) conducted a case report in Coronary artery pseudo-occlusion during percutaneous transluminal coronary angioplasty (PTCA). Active hemoperfusion system during PTCA vs. Termination of active hemoperfusion was evaluated on Angiographic appearance of coronary artery pseudo-occlusion. Active hemoperfusion during PTCA can cause angiographic coronary pseudo-occlusion that resolves with normal vessel opacification upon terminating hemoperfusion.
Synapse has enriched 10 related papers. Consider them for comparative context: