The instrument described below has been in use at a number of hospitals for the past four years.* It was designed as a sub-standard for calibration of X-ray therapy sets. It has, however, found use in other directions, such as the checking of dosage to points on patients and the measurement of H.V.L. and the plotting of depth dose curves. Its accuracy compares favourably with that of other instruments used for dosage measurements, such as the Victoreen dose meter, and it forms a convenient alternative to this instrument for general use in radiotherapy. The dose meter consists essentially of a thimble ionization chamber on the end of a polythene-insulated cable with the inner electrode of the chamber joined directly to the cable core. The other end of the cable is connected to an electrometer valve with the high insulation core joined to the grid, and the anode current of this valve is observed on a microammeter. The valve is contained in a hermetically sealed box to preserve insulation and an “earthing” switch connects its grid to a fixed potential for starting. Saturation in the chamber is produced by connecting a large negative voltage between its case and the valve cathode. Thus, when the chamber is exposed to X rays, the potential of the inner electrode falls and the amount of fall is proportional to the X-ray dose.