High energy radiation has two inherent advantages over low energy radiation in the treatment of deep-seated tumours. These are (a) the high percentage depth dose resulting from the small attenuation of a high energy beam, and (b) the small dose which is given to the skin surface. The design of a therapy unit should be such as to make use of these natural advantages. Methods for achieving a low skin dose are discussed in this paper. Depth dose data for Co60 for a variety of circular fields are presented. The variation of the percentage depth dose with area and with source-to-skin distance is discussed. The effects of the position and thickness of the limiting diaphragm on isodose distributions have been investigated. In the light of this experimental work, suitable treatment cones have been designed. When high energy photons interact with matter, most of the resulting electrons are projected in the forward direction. Thus the maximum dose produced by high energy radiation occurs below the surface, at a depth corresponding to the average range of these electrons. For the same reason, as the thickness of the front wall of an ionization chamber exposed to cobalt γ rays is increased, the response increases until a certain maximum value is obtained, after which the response decreases slowly. The ratio of this maximum response to the response with a very thin front wall is called the build-up ratio.