Does a subcutaneous implantable cardioverter-defibrillator (S-ICD) improve defibrillation success compared to a transvenous ICD in patients with LAMP2 hypertrophic cardiomyopathy?
Does a subcutaneous implantable cardioverter-defibrillator (S-ICD) improve defibrillation success compared to a transvenous ICD in patients with LAMP2 hypertrophic cardiomyopathy?
S-ICD may be more effective than TV-ICD for successful defibrillation in patients with severe hypertrophic cardiomyopathy due to Danon disease.
Answers are grounded in this paper and cite their sources.
S-ICD succeeded where TV-ICD failed in one LAMP2 HCM case; hypothesis-generating and should not change device selection practice.
Danon disease is a rare X-linked lysosomal disease causing severe hypertrophic cardiomyopathy (LAMP2 cardiomyopathy) and an extremely poor prognosis in males, with several reported cases of sudden cardiac death despite the use of transvenous implantable cardioverter defibrillators (TV-ICD). We describe a case in which a TV-ICD was unable to defibrillate induced ventricular fibrillation (VF), but a wholly subcutaneous system (S-ICD) was successful in terminating induced VF and spontaneous ventricular tachycardia. These findings have relevance to the selection of device therapy in the management of these individuals and a wider group of young patients with severe hypertrophic cardiomyopathy.
No takes yet. Share an insight, caveat, or question.
Zaki et al. (2013) studied this question.