Why the study?
Does the double snare technique improve the success rate and efficiency of retrieving embolized ASD or LAA occluders in patients with device embolization?
Does the double snare technique improve the success rate and efficiency of retrieving embolized ASD or LAA occluders in patients with device embolization?
The double snare technique is highly effective and safe for retrieving embolized ASD and LAA occluders, often allowing retrieval through the original delivery sheath.
Answers are grounded in this paper and cite their sources.
May facilitate embolized device retrieval in congenital closures; hypothesis-generating pending larger validation.
BACKGROUND: Device embolization is the most frequent procedural complication during transcatheter closure of congenital cardiac defects. Retrieval of an embolized device may often be complicated by failure to introduce the right atrial (RA) disk hub into the sheath or difficulty in securely grasping the hub pin of RA disk. We aimed to evaluate the efficiency and success rate of device retrieval using a novel double snare technique. METHODS: We reviewed retrieval procedures of embolized atrial septal defect (ASD) or left atrial appendage (LAA) occluder using double snare technique reported from five tertiary referral centers in Korea, Japan, and Uzbekistan. A total of 16 retrieval procedures in 15 patients were reported, including 14 patients who were planned for ASD device closure while 1 patient was planned for LA appendage occlusion. RESULTS: Retrieved devices included 15 ASD occluders from six different manufacturers and one Amplantzer cardiac plug. Success rate of retrieval procedure was 100% using the double snare technique. There were no complications related to device retrieval. Most (15/16, 93.8%) of these devices could be retrieved through their original delivery sheaths. In six patients for whom retrieval was unsuccessful with conventional single snare technique and switched to double snare technique, the retrieval time was shortened significantly (P = 0.004*) by using the double snare technique. CONCLUSIONS: The double snare technique enables effective retrieval of various embolized devices. It abolishes the need of changing the sheath to a larger one in most patients.
No takes yet. Share an insight, caveat, or question.
Ha et al. (2018) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: