Standardization of immunoassays for estradiol-17β (E2) for accuracy was considered in the “Estradiol International Workshops” (1). There, industry asked our laboratory to provide them with a means to evaluate the accuracy/specificity of their E2 test systems. In answer to this demand, we developed and made commercially available an accuracy-based panel of patient specimens. The results of the method comparisons subsequently performed by industry are presented below. Perspectives to use this exercise as a basis for further international standardization efforts are given. The reference panel (tradename: SysRef E2) was prepared through a joint venture between our university laboratory and Scantibodies Laboratory, Inc. (SLI, Santee, CA). To allow a realistic evaluation of routine test systems, the panel consisted of sera from single donations and had E2 concentrations and matrices of clinical relevance: 5 sera were from men and 12 from women. The latter were either in the fertile stage of life, some of them at a defined stage of the menstrual cycle, or postmenopausal. For another five female sera, we deviated from the “single donation” approach by supplementing them with serum from a pregnant woman. This compromise was made because it was considered desirable to include sera with concentrations typical for ovulation-induction therapy, for which it was, unfortunately, not possible to obtain a sufficient volume of blood from such women. The study showed, however, that the samples did behave in the method comparison in an identical manner as the single-donation sera. All blood donations were allowed to clot for 1 h at room temperature and were separated by centrifugation. No preservatives were added. The resulting fresh sera (∼200 mL) were stored for 3 days at 4 °C, then sterile-filtered, placed in 1-mL aliquots in screw-capped EppendorfR cups, frozen, and stored at SLI at −70 °C. All donations were tested for infectious diseases and were found negative for HIV-1/2 antibody, hepatitis C virus antibody, syphilis, and hepatitis B surface antigen. The reference panel was delivered in sets of three 1-mL aliquots per serum. Shipment to the customers was done on solid CO2. At the customer’s laboratory, the sera were to be stored at −20 °C, under which conditions they should be stable for 12 months. We recommended that after thawing at room temperature, the samples should be used within a time period of 8 h and never thawed and frozen several times. The target values were determined at the University of Gent by use of a reference method based on isotope dilution-gas chromatography/mass spectrometry (ID-GC/MS) (2). The measurement protocol consisted of duplicate analyses of each sample on two different occasions. Consistency of measurements was verified by internal accuracy assessment, for which purpose the Certified Reference Materials (CRMs) 577 and 578 from the Community Bureau of Reference (BCR, Brussels, Belgium) were used (3). The mean deviations from the CRM targets were −0.5% (CRM 577; certified concentration, 689 ± 32 pmol/L) and −1.2% (CRM 578; 1340 ± 70 pmol/L). The concentrations of E2 in the 22 sera ranged from 14.67 to 3329 pmol/L. The CVs, based on four measurements, ranged from 0.31% to 2.2%. The E2 routine test systems evaluated in the study (16 in total from 11 manufacturers, see Acknowledgments) comprised manually performed single-analyte immunoassays and automated multi-analyte systems, all direct methods making use of radioactive and nonradioactive labels. The study was conducted on a strictly confidential basis, which means that the identity of the tested immunoassay systems was disclosed only to the scientific organizer of the study. All measurements with the routine test systems were performed in the application laboratory of the participating companies. In most of the cases, the samples were analyzed in replicate (although not always explicitly mentioned), sometimes even with different reagent batches and on different instruments. The mean values were always considered for the comparison. The results obtained were investigated by linear least squares regression analysis (x values, reference method; y values, routine test system). The regression data [slope, intercept, SD of the distribution about the regression line (Sy|xy‖x), and product-moment correlation coefficient (r)] are presented in Table 1 . From these data, the accuracy and specificity of the test systems were judged. For most of the systems, the relationship between the x and y values was considered to be linear from visual inspection of the x/y plots (data not shown). The majority of the systems showed the combination of an intercept >0 pmol/L (11 of 16 systems) and a slope <1 (except for test system F, with a slope of 1.018), reflecting, respectively, over- and undercalibration compared with the reference method in the concentration ranges below and above 600 pmol/L [see insert in Fig. 1 (top) made for the concentration range <600 pmol/L]. Because of the broad concentration range, the values for r (for most systems >0.99) were not considered sensitive enough as a criterion to judge the overall specificity. Instead, specificity was judged from the Sy|x;y‖x values obtained from linear regression. The normalized Sy|x;y‖x values ranging from 26 to 216 pmol/L (after omission of two outliers for test system O) and associated x/y plots revealed considerable gradation in the intrinsic quality of the E2 test systems in terms of overall specificity and showed that some systems needed improvement in this respect. For some of the systems, however, the higher Sy|x;y‖x values were because of a nonlinear relationship between their results and the reference method values (e.g., systems J and P). Linear regression data for the method comparison of 16 E2 routine test systems. There is no relationship between the code of the test systems and the sequence of the companies in the acknowledgment. pmol/L. Linear regression data for the method comparison of 16 E2 routine test systems. There is no relationship between the code of the test systems and the sequence of the companies in the acknowledgment. pmol/L. Results obtained by the E2 routine test systems evaluated vs those by the ID-GC/MS reference method for the SysRef E2 serum panel before (top) and after (bottom) recalibration of each individual set of results on the basis of the split-sample measurements. (Top) Systems C and I were not included, and for systems O and L, respectively, two and one outliers were excluded from the graph. (Bottom) The same systems as above were excluded from the graph, as well as two outliers from system O and one outlier each from systems J, K, L, and P. The broken line represents the line of equality (x = y). Although we are aware that recalibration of a test system on the basis of split-sample measurements should be done only where there is sufficient overall specificity, we were interested in seeing, for this study, the effect of an accuracy-based recalibration. This can be derived from Fig. 1 , representing a combined x/y plot for all evaluated test systems together, respectively, before and after recalibration of the individual sets of results. Recalibration of the test systems was done by use of the respective linear regression equations (see Table 1). For five test systems, it was apparent that the regression equation did not adequately describe the relationship over the entire concentration range. In those cases, we performed linear regression in the three concentration ranges (0–200, 200–600, and 600-3500 pmol/L) and recalibrated the systems by the use of these equations. Fourteen of 16 test systems could be successfully recalibrated by this approach (systems C and I were not included in the graph because of the high Sy|x;y‖x values and also were not recalibrated; for other systems, apparent outlying values were excluded as indicated). From the regression data associated with Fig. 1 , it is obvious that after recalibration, the intermethod compatibility of the measurement results improves to a great extent (Sy|x;y‖x decreases from 205 to 93 pmol/L). If, however, we consider recalibration in the concentration range <600 pmol/L, the effect is less obvious (Sy|x;y‖x decreases only from 53 to 35 pmol/L). This could be indicative of a lack of sensitivity of the test systems at the low end of the working range. Fig. 1 also shows that, in spite of the general improvement of the intermethod compatibility after recalibration, certain samples remained a challenge for some of the test systems (see the samples at 103, 380, and 1620 pmol/L with inter-method CVs of, respectively, 32.8%, 20.0%, and 12.1% vs mean CVs for other samples in the same concentration range of 11.5%, 9.2%, and 5.3%). The reasons for this can be various (unknown interferences, special pathologies, and others). In this context, it should be useful that dedicated panels intended for recalibration of a test system do not contain samples with “special” matrices. On the other hand, panels that are primarily intended for method evaluation should include such samples. In conclusion, the study demonstrated that there is interest by industry for reference panels, as presented here for E2, for the evaluation of the intrinsic quality in terms of accuracy/specificity of their test systems. The method comparison indeed can guide manufacturers in judging whether their routine test system: (a) is suf-ficiently specific and performs accurately enough for its intended application; (b) needs recalibration (e.g., whether a deviation from the reference method values in a defined mathematical way is observed); or (c) needs general improvement of the intrinsic quality (no clear relationship to establish because of high Sy|x;y‖x values). With respect to the use of the panel as a basis for recalibration (case b), international guidelines from scientific organizations such as the International Federation for Clinical Chemistry will be needed concerning the minimum degree of specificity that must be achieved to allow recalibration. Note also that the approach is cost effective because one panel can serve as many as 50 manufacturers, which is more than is presently needed. An additional benefit is that all manufacturers can use the same reference panel, which allows better intercomparison of the different test systems. Also, the workload for the reference laboratory is reduced, which is important with respect to the limited number of reference laboratories performing ID-GC/MS analyses for steroid hormones (4). Last but not least, the efforts undertaken by industry to participate in this study showed that the approach of method comparison by split-sample measurements could lead to a breakthrough in harmonization of routine test systems through accuracy-based standardization. We acknowledge the much appreciated collaboration of the following manufacturers of E2 routine test systems: Abbott Laboratories, Bayer Corp., Biodata S.P.A., bioMérieux, Biosource Europe S.A. (previously Medgenix Diagnostics S.A.), Chiron Diagnostics, Diagnostic Products Corp., Orion Diagnostica, Ortho Clinical Diagnostics, Roche Diagnostic Systems, and Sorin Diagnostics s.r.l. Funding for the ID-GC/MS work was provided by the Belgian Government (contract OOA 12050690) and by the Fund for Medical Scientific Research (contract 33.0007.91).
No takes yet. Share an insight, caveat, or question.
Thienpont et al. (1998) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: