This is the authors' abstract. We don't add key points for this paper.
Wearable smart monitors (WSMs) applied for the estimation of electrophysiological signals are of utmost interest for a non-stressed life. WSM which records heart muscle activities could signalize timely a life-threatening event. The heart muscle activities are typically recorded across the heart at the surface of the body; hence, a WSM monitor requires high-quality surface electrodes. The electrodes used in the clinical settings [i.e. silver/silver chloride (Ag/AgCl) with the gel] are not practical for the daily out of clinic usage. A practical WSM requires the application of a dry electrode with stable and reproducible electrical characteristics. We compared the characteristics of six types of dry electrodes and one gelled electrode during short-term recordings sessions (≈30 s) in real-life conditions: Orbital, monolithic polymer plated with Ag/AgCl, and five rectangular shaped 10 × 6 × 2 mm electrodes (Orbital, Ag electrode, Ag/AgCl electrode, gold electrode and stainless-steel AISI304). The results of a well-controlled analysis which considered motion artifacts, line noise and junction potentials suggest that among the dry electrodes Ag/AgCl performs the best. The Ag/AgCl electrode is in average three times better compared with the stainless-steel electrode often used in WSMs.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Popović‐Maneski et al. (2020) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: