Solid membranes of copper hexacyanoferrate (III) in Areldite are evaluated as thallium (I) sensitive electrode. The membrane electrode gave a linear near Normstian response to thallium (I) ions in the concentration range 10−1 - 5 × 10−4 M and can be used to estimate T1 (I) down to 10−4 M. The responses of the electrode is fast and steady potentials are obtained in less than a minute. The same membrane has been used over a period of six months without any appreciable drift in potential. The electrode can also be used satisfactory in partially non-aqueous media and in presence of a number of interfering ions. It is superior to the existing T1(I) solid membrane electrodes as it can function in alkaline range also.
No takes yet. Share an insight, caveat, or question.
Jain et al. (1982) studied this question.
Synapse has enriched one closely related paper. Consider it for comparative context: