Why the study?
Does magnetic remote catheter ablation reduce the occurrence of junctional rhythm compared to conventional manual ablation in patients with AVNRT?
Does magnetic remote catheter ablation reduce the occurrence of junctional rhythm compared to conventional manual ablation in patients with AVNRT?
Magnetic remote catheter ablation for AVNRT is effective and produces significantly fewer junctional beats than conventional manual ablation, suggesting junctional rhythm is not a mandatory indicator of success with this system.
Answers are grounded in this paper and cite their sources.
Fewer junctional beats with magnetic navigation in AVNRT ablation; hypothesis-generating for junctional rhythm as non-mandatory success marker, needing randomized confirmation.
INTRODUCTION: The occurrence of accelerated junctional rhythm (JR) during radiofrequency ablation of the slow pathway in patients with atrioventricular nodal reentrant tachycardia (AVNRT) is frequent. The aim of the present study was to compare the occurrence of JR during magnetic remote catheter ablation to the conventional manual ablation. METHODS AND RESULTS: Twenty six patients (males: seven; age: 51 + or - 15 years) underwent slow pathway ablation with magnetic navigation (MN) system (Niobe, Stereotaxis Inc., St. Louis, MO, USA) and were compared to a control group of 11 patients (males: three; age: 53 + or - 16 years) treated with conventional manual ablation. A 4-mm nonirrigated tip catheter was used in both groups with a maximum of 30 W and 60 degrees C. Acute success was obtained in all patients. In the MN group, three patients out of 24 had no junctional beat (JB) at all and seven patients had 10 or less JB. In contrast, in the conventional group no patient had less than 10 JB. The mean number of JB in the MN group was 66 + or - 94.9 (0-410) and 200 + or - 243.1 (43-914) in the control group (P = 0.019). In the MN group one patient had a first-degree atrioventricular block. No other complication occurred. CONCLUSIONS: Magnetic remote catheter ablation of AVNRT is effective and is associated with less JB than the manual conventional technique. Therefore, JB may not be considered as a mandatory indicator for successful AVNRT ablation with MN system.
No takes yet. Share an insight, caveat, or question.
Ricard et al. (2009) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: