Conditions for efficient and reproducible liquid scintillation counting of C14 using a single phototube are described. The amplifier, scaler and power supply are the same as those used for the gas-phase proportional counting of C14O2 admixed with P-10 gas (9:1 argon-methane). A compound which will dissolve to the extent of about 10 mg/25 ml of xylene or 4:1 xylene-ethanol, which does not absorb light in the blue region and which does not quench phosphor scintillations can be assayed for C14 by this method with a sensitivity about equivalent to that of the gas-phase counting method which entails combustion of the compound to CO2. The counting efficiency in xylene is about 80% and in 4:1 xylene-ethanol about 70% of the C14 disintegrations occurring in the sample. The background counting rate is about 90 counts per minute. Glucose, isolated as the phenylosazone from a biological fluid, is converted to the glucosotriazole derivative. The derivative is held in solution, with the aid of H3BO3, in 25 ml of 4:1 xylene-ethanol containing 0.24% of the phosphor PBD (2-phenyl-5-(4 biphenylyl) 1,3,4-oxadizole). The standard deviation for individual samples in a specific activity determination, including the error in weighing the 10-mg sample of glucosotriazole derivative, is ±1.7% Submitted on July 2, 1956
No takes yet. Share an insight, caveat, or question.
Steele et al. (1957) studied this question.