Why the study?
The study was conducted to evaluate variables that affect the risk of contamination for endoscopic retrograde cholangiopancreatography and endoscopic ultrasound endoscopes.
Does double high-level disinfection or disposable tips reduce the risk of endoscope contamination compared to HLD-ETO or standard tips in ERCP and EUS endoscopes?
Does double high-level disinfection or disposable tips reduce the risk of endoscope contamination compared to HLD-ETO or standard tips in ERCP and EUS endoscopes?
There was no significant reduction in endoscope contamination using HLD-ETO versus dHLD reprocessing, though disposable tips may reduce duodenoscope contamination risk.
Linear echoendoscopes may carry higher contamination risk than duodenoscopes after reprocessing; leaves open optimal protocols for these scopes.
OBJECTIVE: To evaluate variables that affect risk of contamination for endoscopic retrograde cholangiopancreatography and endoscopic ultrasound endoscopes. DESIGN: Observational, quality improvement study. SETTING: University medical center with a gastrointestinal endoscopy service performing ∼1,000 endoscopic retrograde cholangiopancreatography and ∼1,000 endoscopic ultrasound endoscope procedures annually. METHODS: Duodenoscope and linear echoendoscope sampling (from the elevator mechanism and instrument channel) was performed from June 2020 through September 2021. Operational changes during this period included standard reprocessing with high-level disinfection with ethylene oxide gas sterilization (HLD-ETO) was switched to double high-level disinfection (dHLD) (June 16, 2020-July 15, 2020), and duodenoscopes changed to disposable tip model (March 2021). The frequency of contamination for the co-primary outcomes were characterized by calculated risk ratios. RESULTS: The overall pathogenic contamination rate was 4.72% (6 of 127). Compared to duodenoscopes, linear echoendoscopes had a contamination risk ratio of 3.64 (95% confidence interval [CI], 0.69-19.1). Reprocessing using HLD-ETO was associated with a contamination risk ratio of 0.29 (95% CI, 0.06-1.54). Linear echoendoscopes undergoing dHLD had the highest risk of contamination (2 of 18, 11.1%), and duodenoscopes undergoing HLD-ETO and the lowest risk of contamination (0 of 53, 0%). Duodenoscopes with a disposable tip had a 0% contamination rate (0 of 27). CONCLUSIONS: We did not detect a significant reduction in endoscope contamination using HLD-ETO versus dHLD reprocessing. Linear echoendoscopes have a risk of contamination similar to that of duodenoscopes. Disposable tips may reduce the risk of duodenoscope contamination.
No takes yet. Share an insight, caveat, or question.
Ayres et al. (2023) studied this question.