Quantitative gait analysis enables clinicians to evaluate patient mobility and to diagnose neuromuscular disorders. The clinical application of gait analysis is currently limited by the high operating costs of gait laboratories. The use of instrumented footwear that performs out of the lab measurements on subjects' walking patterns is a promising way to overcome this limitation. Besides serving as assessment tools, such devices can also act as retraining tools that help regulate a patient's gait with acoustic or vibrotactile stimuli. At the Columbia University Robotics and Rehabilitation (ROAR) Lab, we have developed SoleSound, a fully portable instrumented footwear that can measure spatiotemporal gait parameters and deliver action-related audio-tactile feedback. This paper presents the validation of SoleSound as a tool for quantitative gait analysis. Using a reference system, we assessed the performance of SoleSound under two calibration strategies, subject-specific and generic. With the latter calibration, mean accuracy and precision (RMSE ± SD) were 2.93 ± 1.32 cm for stride length, 0.70 ± 0.37 cm for foot-ground clearance, 1.54 ± 0.70 cm for step width, 4.53 ± 0.90 cm for foot trajectory and 4.33 ± 1.01 deg for ankle plantar-dorsiflexion angle. Accuracy was further improved when subject-specific calibration was used. These results indicate that SoleSound has the potential to be used for out-of-lab gait analysis.
No takes yet. Share an insight, caveat, or question.
Minto et al. (2015) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: