Purpose Evaluate the use of both single‐echo gradient recalled echo (SE‐GRE) and EPI approaches to creating temperature maps on a mid‐field head‐only scanner, both in vivo and on a tissue mimicking gel. Methods Three 2D protocols were investigated (an SE‐GRE, single‐shot EPI, and an averaged single‐shot EPI). The protocols used either a gradient recalled acquisition or an echo planar acquisition, with EPI parameters optimized for the longer at lower field‐strengths. Phantom experiments were conducted to evaluate temperature tracking while cooling, comparing protocol to measurements from an optical fiber thermometer. Studies were performed on a 0.5T head only MR scanner. Temperature stability maps were produced in vivo for the various protocols to evaluate precision. Results The use of an EPI protocol for thermometry improved temperature precision in a temperature control phantom and provided an 18% improvement in temperature measurement precision in vivo. Temperature tracking using a fast (<2 s) update rate EPI thermometry sequence provided a similar precision to the slower SE‐GRE protocol. Conclusion While SE‐GRE PRF thermometry shows good performance, EPI methods offer improved tracking precision or update rate, making them a better option for thermometry in the brain at mid‐field.