Why the study?
Does measuring ST deviation at 10 ms after the J point improve diagnostic accuracy for acute ischemia in patients with suspected acute coronary syndrome?
Does measuring ST deviation at 10 ms after the J point improve diagnostic accuracy for acute ischemia in patients with suspected acute coronary syndrome?
Measuring ST deviation close to the J point (specifically at 10 ms after) using a liberal detection algorithm optimizes the diagnostic accuracy for acute ischemia on ECG.
Supports measuring ST deviation 10 ms after J point for ischemia detection; leaves open prospective validation before practice change.
BACKGROUND: There is no consensus about the time instant relative to the J point where ST deviation has to be measured for detection of acute ischemia in the ECG. METHODS: We analyzed 53 ECGs, recorded preceding emergency catheterization of acute coronary syndrome patients with a completely occluded culprit artery (cases), and 88 control ECGs recorded in the cardiology outpatient clinic. ECG-amplitude measurements were made every 10 ms, between 20 ms before till 80 ms after the J point. STEMI-detection algorithms varied from the traditional STEMI criterion (elevations in at least two adjacent ECG leads), via the STEMI equivalent criterion (depressions in V2 and V3), to the most liberal STEMI-detection algorithm in which elevations as well as depressions in two adjacent leads were considered as signs of ischemia. RESULTS: Diagnostic accuracy was highest (93.6%) for the most liberal STEMI-detection algorithm at 10 ms after the J point; sensitivity was 94.3% and specificity was 93.2%. CONCLUSION: The results of our study suggest that STEMI detection close to the J point is optimal.
No takes yet. Share an insight, caveat, or question.
Man et al. (2016) studied this question.