Can laser Doppler velocimetry be used to measure blood velocity in human retinal vessels?
Can laser Doppler velocimetry be used to measure blood velocity in human retinal vessels?
Laser Doppler velocimetry can successfully measure blood flow velocities in human retinal arteries and veins.
Answers are grounded in this paper and cite their sources.
Feasibility of retinal velocimetry shown; leaves open clinical utility and validation in larger cohorts.
Laser Doppler velocimetry was used to measure the velocity of blood in human retinal vessels. The mean flow velocities obtained were 1.9 centimeters per second in a retinal vein and 2.2 centimeters per second in a retinal artery. Scattered light from a weak helium-neon laser beam focused on the vessel was detected by a photomultiplier, and the temporal correlation of the intensity fluctuations was measured with a photon counting autocorrelator. Autocorrelation functions for blood flowing through glass capillaries were used for calibration.
No takes yet. Share an insight, caveat, or question.
Tanaka et al. (1974) studied this question.