Transimpedance amplifiers are current-to-voltage converters. They are found in a wide variety of applications that require current measurements at high sensitivity, low noise and/or high bandwidth. Inherent to the construction of such an amplifier are certain tradeoffs between bandwidth, amplification, and noise level, requiring a careful design and layout of a transimpedance amplifier for any given task. These design limitations become especially relevant for high-gain, high-bandwidth applications, such as photodiode readout or ion current sensing. In these applications, commercially available amplifiers are often not suited for the task, leading to a low signal-to-noise ratio. This paper discusses the most significant limiting factors of transimpedance amplifiers, and describes their effects, source and possible mitigation options. It presents a novel combination of feedback resistor configuration and shielding layout to improve performance beyond typical single-resistor shielded options. Based on this, a low-cost transimpedance amplifier made with readily available off-the-shelf components reaching 30 kHz bandwidth and 1 GΩ amplification with only 3.898 pA root-mean-square noise was developed, characterized and demonstrated for use in ion mobility spectrometry as one exemplary application requiring high gain, low noise and high bandwidth.
No takes yet. Share an insight, caveat, or question.
Bailey et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: