Why the study?
Does percutaneous mitral valve repair using the MitraClip system reduce mitral valve annular size and improve coaptation in patients with mitral regurgitation?
Does percutaneous mitral valve repair using the MitraClip system reduce mitral valve annular size and improve coaptation in patients with mitral regurgitation?
PMVR using the MitraClip system mechanically approximates mitral valve annulus edges, reducing annular size and improving coaptation, which correlates with less residual mitral regurgitation.
PMVR reduces mitral annular dimensions alongside MR improvement; leaves open whether annular remodeling independently drives outcomes in prospective trials.
Aims: Improved mitral valve leaflet coaptation with consecutive reduction of mitral regurgitation (MR) is a central goal of percutaneous mitral valve repair (PMVR) with the MitraClip® system. As influences of PMVR on mitral valve geometry have been suggested before, we examined the effect of the procedure on mitral annular size in relation to procedural outcome. Methods and results: Geometry of the mitral valve annulus was evaluated in 183 patients undergoing PMVR using echocardiography before and after the procedure and at follow-up. Mitral valve annular anterior-posterior (ap) diameter decreased from 34.0 ± 4.3 to 31.3 ± 4.9 mm (P < 0.001), and medio-lateral (ml) diameter from 33.2 ± 4.8 to 32.4 ± 4.9 mm (P < 0.001). Accordingly, we observed an increase in MV leaflet coaptation after PMVR. The reduction of mitral valve ap diameter showed a significant inverse correlation with residual MR. Importantly, the reduction of mitral valve ap diameter persisted at follow-up (31.3 ± 4.9 mm post PMVR, 28.4 ± 5.3 mm at follow-up). Conclusion: This study demonstrates mechanical approximation of both mitral valve annulus edges with improved mitral valve annular coaptation by PMVR using the MitraClip® system, which correlates with residual MR in patients with MR.
No takes yet. Share an insight, caveat, or question.
Patzelt et al. (2017) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: