Answers are grounded in this paper and cite their sources.
The handling of low, middle and high molecular weight markers was examined in seven stable dialysis patients during hemofiltration with different membranes. Four membranes were examined in a randomized, crossover order (polysulfone, polyamide, AN69 polyacrylonitrile, Asahi polyacrylonitrile) by measuring plasma and dialysate concentrations of phosphate, creatinine, vitamin B12, beta 2-microglobulin, furanic acid, hippuric acid, retinol-binding protein, alpha-1-antitrypsin, and albumin. Sieving coefficients and plasma clearances of beta 2-microglobulin or retinol-binding protein were markedly or slightly lower during hemofiltration with the Asahi polyacrylonitrile membrane than with the other membranes (highest removal with polysulfone/AN69 polyacrylonitrile membranes). No differences of obvious clinical relevance could be seen between the four membranes. A high beta 2-microglobulin removal rate might be important to prevent dialysis-associated amyloidosis.
No takes yet. Share an insight, caveat, or question.
Krämer et al. (1992) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: