Answers are grounded in this paper and cite their sources.
OBJECTIVE: We propose a novel attachment method for retinal tissue that utilizes silicone modified with bioactive molecules. DESIGN: section performed in rabbits. SUBJECTS: experiments 4 rabbits were used. METHODS: was implanted into a rabbit's eye after a full vitrectomy was performed. Post-operative exams were done every two weeks to evaluate placement, attachment and sterilization method. After three months animals were euthanized and eye was enucleated for histology analysis. MAIN OUTCOME MEASURES: studies for biocompatibility purposes. RESULTS: experiments. A sterilization protocol was tested and proved to be reliable for bioactive materials. Disintegrin coated silicone showed attachment in 2 of 4 rabbits during the 3-month implant period. The adhesion was persistent until reversed with plasmin. All rabbits were implanted for 3 months regardless of attachment, to test the feasibility of the sterilization method. None of the rabbits developed any type of eye infection during the implant period. CONCLUSION: We successfully lased and produced an active area on the silicone surface to allow disintegrin-silicone binding. Different silicones interact differently with the laser energy, and this is reflected in the strength of the silicone-disintegrin-retina attachment.