Key result
The Philips health watch accurately estimated total energy expenditure within a 15% equivalence margin (MAPE 10%) and resting heart rate within a 10% margin compared to medical reference devices.
Why the study?
Does the Philips health watch accurately measure total energy expenditure and resting heart rate compared to medical reference devices in healthy adults?
Cross-Sectional (n=29)
No
Does the Philips health watch accurately measure total energy expenditure and resting heart rate compared to medical reference devices in healthy adults?
Effect estimate: 97% ratio (95% CI 92-101)
p-value: p=<0.05
The Philips health watch accurately measures resting heart rate and total energy expenditure, providing a reliable tool for unobtrusive monitoring in lifestyle interventions.
Supports wearable TEE monitoring in research; leaves open clinical adoption without further validation.
BACKGROUND: Physical inactivity is an important modifiable risk factor for chronic diseases. A new wrist-worn heart rate and activity monitor has been developed for unobtrusive data collection to aid prevention and management of lifestyle-related chronic diseases by means of behavioral change programs. OBJECTIVE: The objective of the study was to evaluate the performance of total energy expenditure and resting heart rate measures of the Philips health watch. Secondary objectives included the assessment of accuracy of other output parameters of the monitor: heart rate, respiration rate at rest, step count, and activity type recognition. METHODS: A within-person comparative study was performed to assess the performance of the health watch against (medical) reference measures. Participants executed a protocol including 15 minutes of rest and various activities of daily life. A two one-sided tests approach was adopted for testing equivalence. In addition, error metrics such as mean error and mean absolute percentage error (MAPE) were calculated. RESULTS: ) completed the 81-minute protocol. Their mean resting heart rate in beats per minute (bpm) was 64 (SD 7.3). With a mean error of -10 (SD 38.9) kcal and a MAPE of 10% (SD 8.7%), total energy expenditure estimation of the health watch was found to be within the 15% predefined equivalence margin in reference to a portable indirect calorimeter. Resting heart rate determined during a 15-minute rest protocol was found to be within a 10% equivalence margin in reference to a wearable electrocardiogram (ECG) monitor, with a mean deviation of 0 bpm and a maximum deviation of 3 bpm. Heart rate was within 10 bpm and 10% of the ECG monitor reference for 93% of the duration of the protocol. Step count estimates were on average 21 counts lower than a waist-mounted step counter over all walking activities combined, with a MAPE of 3.5% (SD 2.4%). Resting respiration rate was on average 0.7 (SD 1.1) breaths per minute lower than the reference measurement by the spirometer embedded in the indirect calorimeter during the 15-minute rest, resulting in a MAPE of 8.3% (SD 7.0%). Activity type recognition of walking, running, cycling, or other was overall 90% accurate in reference to the activities performed. CONCLUSIONS: The health watch can serve its medical purpose of measuring resting heart rate and total energy expenditure over time in an unobtrusive manner, thereby providing valuable data for the prevention and management of lifestyle-related chronic diseases. TRIAL REGISTRATION: Netherlands trial register NTR5552; http://www.trialregister.nl/trialreg/admin/rctview.asp?TC=5552 (Archived by WebCite at http://www.webcitation.org/6neYJgysl).
No takes yet. Share an insight, caveat, or question.
Hendrikx et al. (2017) conducted a cross-sectional in Healthy adults (n=29). Philips health watch vs. COSMED K4b2 and Actiwave Cardio reference devices was evaluated on Total energy expenditure estimation accuracy (ratio of health watch to reference) (97% ratio, 95% CI 92-101, p=<0.05). The Philips health watch accurately estimated total energy expenditure within a 15% equivalence margin (MAPE 10%) and resting heart rate within a 10% margin compared to medical reference devices.