Key result
A deep learning model automatically measured left ventricular outflow tract diameter with a median absolute difference of 0.97 mm (95% CI 0.79-1.14) compared to cardiologists across all image qualities.
Why the study?
Significant inter- and intraobserver variability in LVOTd measurement impacts aortic stenosis evaluation and cardiac output calculation due to squaring of the radius.
Does a deep learning model accurately measure LVOTd in 2D echocardiography compared to cardiologists?
Observational (n=656)
No
Does a deep learning model accurately measure LVOTd in 2D echocardiography compared to cardiologists?
Effect estimate: Median absolute difference 0.97 mm (95% CI 0.79-1.14)
A deep learning model can automatically measure left ventricular outflow tract diameter on 2D echocardiography with accuracy comparable to cardiologists, potentially improving measurement consistency.
May support automated echocardiography analysis; leaves open prospective clinical validation before routine use.
Background Left ventricular outflow tract diameter (LVOTd) is routinely measured to calculate stroke volume and estimate aortic valve area by the continuity equation. Despite LVOTd being regularly measured clinically, significant inter- and intraobserver variability is evident. This variability is highly impactful on both aortic stenosis evaluation and cardiac output calculation due to the squaring of the LVOT radius. Purpose We aimed to investigate if LVOTd measurements from clinical echocardiographic examinations could be used in a deep learning (DL) model to automatically perform LVOTd measurements with equivalent accuracy and improved consistency compared to current practice. Methods Data was collected from clinical echocardiographic examinations performed on 656 consecutive patients admitted to the cardiac catheterization laboratory at a university hospital in January – December 2018. Parasternal views with cardiologist annotated LVOTd coordinates were assessed for 1314 echocardiographic still images. The quality of the still image and annotated LVOT ground truth were individually graded as high, medium and low by experienced cardiologists to establish a rigorous training basis. Spatial geometry data was preserved for each still image in order to distinguish between different degrees of image zoom. Data was randomly split into training, validation and testing sets (68%, 17%, 15%). A fully convolutional network based on the Resnet50 architecture was used with a custom loss function with heatmap regression. Image augmentations were added to extend the dataset. Results When including echocardiographic images of any quality (n=1314) in the model training and inference, the median absolute difference between cardiologist LVOTd and DL LVOTd was 0.97 mm (95% Confidence interval (CI) 0.79–1.14). Using only high and medium quality still images and ground truth (n=869) in the training and inference, median absolute difference decreased to 0.81 mm (95% CI 0.60–0.96). Adding image augmentations to this dataset further improved the model, resulting in a median LVOTd absolute difference of 0.66 mm (95% CI 0.51–0.78). The LVOTd error in inference increased with decreasing image quality, as shown in Figure 1, with two predictions (0.9%) failing completely. Conclusion Deep learning models are capable of measuring LVOTd with comparable accuracy to cardiologists when trained on clinical data. Data quality affects both training and inference. Even with a slightly lower accuracy when used on lower quality echocardiographic images, DL-assisted LVOT measurement has a clear potential to increase repeatability and consistency of LVOTd measurements. Funding Acknowledgement Type of funding sources: Public Institution(s). Main funding source(s): Research Council of Norway (Norges forskningsråd) Figure 1
No takes yet. Share an insight, caveat, or question.
ZHA et al. (2021) conducted an observational in Patients undergoing clinical echocardiographic examinations (n=656). Deep learning model for automated LVOTd measurement vs. Cardiologist annotated LVOTd was evaluated on Median absolute difference between cardiologist LVOTd and deep learning LVOTd (all images) (Median absolute difference 0.97 mm, 95% CI 0.79-1.14). A deep learning model automatically measured left ventricular outflow tract diameter with a median absolute difference of 0.97 mm (95% CI 0.79-1.14) compared to cardiologists across all image qualities.