Key result
Automated signal processing demonstrated that atrial near-field and ventricular far-field signal amplitudes and morphologies did not significantly change during exercise or change of posture.
Observational (n=23)
No
Automated signal processing can reliably classify atrial near and far fields, demonstrating that signal morphology remains stable during exercise and posture changes, which is a prerequisite for beat-to-beat template analysis in pacemakers.
Answers are grounded in this paper and cite their sources.
Signal stability supports template-based algorithms in devices; leaves open prospective validation for arrhythmia discrimination.
INTRODUCTION: Sophisticated monitoring of atrial activity is a prerequisite for modern pacemaker therapy. Ideally, near-fields and ventricular far-fields ought to be distinguished by beat-to-beat template analysis of the atrial signal. A prerequisite is that atrial signals are stable under different conditions. METHODS AND RESULTS: A Matlab routine was developed to analyze atrial electrograms of 23 patients at least 3 months after implantation of a dual chamber pacemaker under several conditions including at rest, bipolar at rest, in an upright position, during treadmill exercise, and postexercise. A near-field and far-field template was created and amplitudes, widths, and slew rates were measured. In bipolar configuration, near-field amplitude at rest was 3.04 +/- 0.94 mV (unipolar)/3.36 +/- 1.0 mV (bipolar) versus 3.18 +/- 1.0 mV (bipolar) at peak exercise. Far-field amplitude at rest was 1.66 +/- 1.18 (unipolar)/0.47 +/- 0.27 mV (bipolar) and 0.41 +/- 0.21 mV (bipolar) at peak exercise (n.s. for bipolar measurements). No overall significant changes were observed for near- and far-field widths and slew rates during exercise. Shorter tip-ring distances of the atrial bipole, lead position, and the presence of sinus node disease did not have any impact on overall near- and far-field signal characteristics. Intraindividual differences between rest and peak exercise were moderate (range: near-field +0.15 to -0.54 mV; range: far-field +0.05 to -0.18 mV). CONCLUSIONS: Atrial near and far fields can be automatically classified and quantified by automated signal processing. Signals did not change during exercise or change of posture. This is a prerequisite for the implementation of beat-to-beat template analysis into pacemakers.
No takes yet. Share an insight, caveat, or question.
Eberhardt et al. (2006) conducted an observational in Patients with dual chamber pacemakers (n=23). Exercise and posture changes vs. Rest was evaluated on Bipolar near-field amplitude (mV) (p=n.s.). Automated signal processing demonstrated that atrial near-field and ventricular far-field signal amplitudes and morphologies did not significantly change during exercise or change of posture.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: