Key result
Far-field ventricular signals show smaller amplitude and slew rate than atrial deflections during pacemaker implantation.
Observational (n=200)
Accurate sensing of far-field ventricular signals from an atrial pacing lead is feasible in most patients, which may help improve physiological atrial rate responsive and DDD(R) pacing algorithms.
Answers are grounded in this paper and cite their sources.
Feasibility of far-field ventricular sensing from atrial leads may aid DDD(R) algorithms; leaves open prospective validation before practice change.
Accurate detection of the spontaneous far-field ventricular signal may be used to determine the ventricular activation, and hence, the interval from atrial stimulus to the ventricular R wave (AR interval) using a standard atrial pacing lead. This can be useful in developing a physiological atrial rate responsive (AAIR) pacemaker and in further improving DDD(R) pacing algorithms. In order to better characterize the atrial sensed far-field ventricular signal, 200 consecutive patients undergoing pacemaker implantation were studied. The amplitude of the far-field ventricular signal was significantly smaller than that of the atrial deflection. In all recordings, the slew rate of the atrial deflection was larger than that of the far-field ventricular signal. Subdivision of the recordings by electrode position, pocket location, or QRS duration on the surface ECG resulted in significantly different signal characteristics. The amplitude and slew rate of the far-field ventricular signal were significantly smaller in bipolar versus unipolar sensing. Atrial sensed far-field ventricular recordings could also be obtained in the case of ventricular pacing. Our results indicate that accurate sensing of the far-field ventricular signal from an atrial pacing lead is conceivable in most patients. The different signal characteristics in relation to parameters, such as electrode position, sensing mode, and pocket location, may be useful in determining the optimal conditions for signal sensing.
No takes yet. Share an insight, caveat, or question.
Brouwer et al. (1997) conducted an observational in Pacemaker implantation (n=200). Atrial sensed far-field ventricular signal vs. Atrial deflection was evaluated on Amplitude and slew rate of the far-field ventricular signal. The amplitude and slew rate of the far-field ventricular signal were significantly smaller than those of the atrial deflection in patients undergoing pacemaker implantation.
Synapse has enriched 10 related papers. Consider them for comparative context: