Key result
Intravascular MRI-guided high-intensity focused ultrasound ablation successfully produced targeted thermal lesions in ex vivo tissues and 2 live pigs, with MRI thermometry highly correlating with histology (r ≥ 0.79).
Why the study?
To develop and test an intravascular MRI-guided high-intensity focused ultrasound ablation method for targeting perivascular pathology with minimal vessel wall injury.
Effect estimate: r ≥ 0.79
Intravascular MRI-guided HIFU is a feasible method for creating controlled perivascular ablations while preserving the vessel wall in preclinical models.
Hypothesis-generating for intravascular MRI-guided HIFU ablation; leaves open clinical translation pending rigorous in vivo validation.
Purpose To develop and test in animal studies ex vivo and in vivo, an intravascular (IV) MRI‐guided high‐intensity focused ultrasound (HIFU) ablation method for targeting perivascular pathology with minimal injury to the vessel wall. Methods IV‐MRI antennas were combined with 2‐ to 4‐mm diameter water‐cooled IV‐ultrasound ablation catheters for IV‐MRI on a 3T clinical MRI scanner. A software interface was developed for monitoring thermal dose with real‐time MRI thermometry, and an MRI‐guided ablation protocol developed by repeat testing on muscle and liver tissue ex vivo. MRI thermal dose was measured as cumulative equivalent minutes at 43°C (CEM 43 ). The IV‐MRI IV‐HIFU protocol was then tested by targeting perivascular ablations from the inferior vena cava of 2 pigs in vivo. Thermal dose and lesions were compared by gross and histological examination. Results Ex vivo experiments yielded a 6‐min ablation protocol with the IV‐ultrasound catheter coolant at 3‐4°C, a 30 mL/min flow rate, and 7 W ablation power. In 8 experiments, 5‐ to 10‐mm thick thermal lesions of area 0.5‐2 cm 2 were produced that spared 1‐ to 2‐mm margins of tissue abutting the catheters. The radial depths, areas, and preserved margins of ablation lesions measured from gross histology were highly correlated (r ≥ 0.79) with those measured from the CEM 43 = 340 necrosis threshold determined by MRI thermometry. The psoas muscle was successfully targeted in the 2 live pigs, with the resulting ablations controlled under IV‐MRI guidance. Conclusion IV‐MRI‐guided, IV‐HIFU has potential as a precision treatment option that could preserve critical blood vessel wall during ablation of nonresectable perivascular tumors or other pathologies.
No takes yet. Share an insight, caveat, or question.
Liu et al. (2019) studied Perivascular pathology (n=2). Intravascular MRI-guided high-intensity focused ultrasound (IV-HIFU) ablation was evaluated on Thermal dose and lesions compared by gross and histological examination (r ≥ 0.79). Intravascular MRI-guided high-intensity focused ultrasound ablation successfully produced targeted thermal lesions in ex vivo tissues and 2 live pigs, with MRI thermometry highly correlating with histology (r ≥ 0.79).