Key result
Transendocardial ethanol injection using a novel intracardiac echocardiography catheter successfully guided needle position and resulted in controlled myocardial infarction in 12 canines.
Why the study?
Does a novel multifunctional intracardiac echocardiography catheter enable feasible and safe target septal ablation with transendocardial ethanol injection in canines?
Does a novel multifunctional intracardiac echocardiography catheter enable feasible and safe target septal ablation with transendocardial ethanol injection in canines?
A novel intracardiac echocardiography catheter enables visually guided transendocardial ethanol injection for controlled septal ablation in a canine model, offering a potential new approach for treating hypertrophic obstructive cardiomyopathy.
Feasible in a canine model; leaves open translation to safe septal ablation in hypertrophic obstructive cardiomyopathy.
OBJECTIVES: To evaluate the feasibility and safety of a novel multifunctional intracardiac echocardiography catheter for target septal ablation with transendocardial ethanol injection in canines. BACKGROUND: Percutaneous transluminal septal myocardial ablation has been the primary catheter-based strategy for the treatment of hypertrophic obstructive cardiomyopathy (HOCM). However, inherent limitations of the technique have reduced its therapeutic efficacy. METHODS: A new catheter (10 F), integrated with a nitinol needle (29 G) and a 6.5-10-MHz, 32-element, side-fire ultrasound imaging transducer, was delivered into the right ventricle in nine canines and the left ventricle in three canines. A 0.3-ml microbubble and 0.5-1.5 ml absolute ethanol were sequentially injected into the interventricular septum. Electrocardiograph recordings were obtained during the whole procedure. Three hours after this operation, the heart was harvested for gross and histological examination. RESULTS: In all canines, the catheter provided the structural support and helped guide proper needle position within the septum. The microbubble further allowed the confirmation of the needle location through focal echo-density enhancement. Different amounts of ethanol infusion produced a dose-related effect on myocardial ablation. Macroscopic examination showed that the target myocardium became pale with a distinct border between lesions and normal tissue. Hematoxylin and eosin staining further confirmed necrosis within the injection region. CONCLUSIONS: Transendocardial ethanol injection at the interventricular septum resulted in controlled myocardial infarction. In addition, the injection depth could be visually followed using this new system, which may provide a safer and more intuitive method for the treatment of HOCM or other cardiomyopathies.
No takes yet. Share an insight, caveat, or question.
Guo et al. (2011) studied Hypertrophic obstructive cardiomyopathy (HOCM) (n=12). Transendocardial ethanol injection using a multifunctional intracardiac echocardiography catheter was evaluated on Feasibility and safety of target septal ablation. Transendocardial ethanol injection using a novel intracardiac echocardiography catheter successfully guided needle position and resulted in controlled myocardial infarction in 12 canines.