A large parallax-free gas diffraction meter based on a thinner-THGEM (thick gaseous electron multiplier) has been developed at the Beijing Synchrotron Radiation Facility (BSRF). A thinner-THGEM of thickness 200 µm is adopted, which can be shaped into a curve to eliminate parallax-error effects. The detector is designed to have a 48° open angle positioned 20 cm from the powder samples. A front-end electronics board with 128 channels direct-current mode was adapted for the 8 keV BSRF beamline with 0.2 ns/100 ns stable duty cycle. Two powder samples, TiO 2 and SnO 2 , were tested separately. The measured spectra with an angular resolution of 0.148 ± 0.081° are consistent with the data from the powder diffraction file. Combining the gas gain of the thinner-THGEM with the electronic circuit dynamic range, a very broad dynamic range of about 10 7 could be obtained.