Why the study?
Does optimizing image acquisition and signal processing parameters improve the precision and reliability of pulse wave imaging (PWI) in ex vivo canine aortas and in vivo human arteries?
Does optimizing image acquisition and signal processing parameters improve the precision and reliability of pulse wave imaging (PWI) in ex vivo canine aortas and in vivo human arteries?
Optimizing spatial resolution (24-48 scan lines) and tracking the 50% upstroke of the waveform improves the precision and reliability of pulse wave imaging for measuring pulse wave velocity using conventional ultrasound.
Answers are grounded in this paper and cite their sources.
Optimizing PWI parameters may improve research measurement precision; leaves open clinical adoption and outcome impact.
The propagation behavior of the arterial pulse wave may provide valuable diagnostic information for cardiovascular pathology.pulse wave imaging (PWI) is a noninvasive, ultrasound imaging-based technique capable of mapping multiple wall motion waveforms along a short arterial segment over a single cardiac cycle, allowing for the regional pulse wave velocity (PWV) and propagation uniformity to be evaluated.The purpose of this study was to improve the clinical utility of PWI using a conventional ultrasound system.The tradeoff between PWI spatial and temporal resolution was evaluated using an ex vivo canine aorta (n Z 2) setup to assess the effects of varying image acquisition and signal processing parameters on the measurement of the PWV and the pulse wave propagation uniformity r 2 .PWI was also performed on the carotid arteries and abdominal aortas of 10 healthy volunteers (24.8 AE 3.3 y.o.) to determine the waveform tracking feature that would yield the most precise PWV measurements and highest r 2 values in vivo.The ex vivo results indicated that the highest precision for measuring PWVs w2.5e3.5 m/s was achieved using 24e48 scan lines within a 38 mm image plane width (i.e.0.63e1.26lines/mm).The in vivo results indicated that tracking the 50% upstroke of the waveform would consistently yield the most precise PWV measurements and minimize the error in the propagation uniformity measurement.Such findings may help establish the optimal image acquisition and signal processing parameters that may improve the reliability of PWI as a clinical measurement tool.
No takes yet. Share an insight, caveat, or question.
Li et al. (2015) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: