Dynamic MRI with high temporal resolution combined with a cross-correlation algorithm provides a reproducible, noninvasive method for estimating local carotid pulse wave velocity.
May support reproducible noninvasive carotid PWV via MRI cross-correlation; leaves open outcome validation before clinical adoption.
The current study presents a new protocol for local pulse wave velocity (PWV) measurement using dynamic MR sequences, which have a high temporal resolution (TR < 6 ms). MR images were obtained at two positions along the common carotid artery, separated by a distance of 5 cm. In each phase of a MR series, carotid region was automatically extracted and then its area distension waveform could be obtained. Sixteen volunteers with no symptoms of cardiovascular diseases were studied. For local PWV estimation, three delay estimation principles were tested and produced the following values: intersecting tangents method (M1): 4.72 ± 1.40 m/s, second derivative method (M2): 4.94 ± 1.68 m/s and cross-correlation method (M3): 5.03 ± 1.17 m/s. The cross-correlation method showed a relative high reliability as its least standard deviation.