Answers are grounded in this paper and cite their sources.
A wireless energy transmitting system is required for patient-implanted ventricular assist devices. We developed a wireless power feeding system that uses a large transmitting coil embedded in the floor to send power to a receiving coil implanted between the skin and fat of the human body. To obtain higher transmission efficiency, we examined three types of transmitting coil models (Models 1-3) in this study. The number of turns N and winding interval t of Models 1, 2, and 3 were N = 20, t = 12 mm; N = 10, t = 25.6 mm; and N = 6, t = 46 mm, respectively. We investigated the tradeoff between the self-resonance frequency of the transmitting coil and mutual inductance of the transformer. The results showed that a maximum transmission efficiency of 39.7% (standing position) and 75.8% (seating position) was obtained at 1 MHz with Model 2. Furthermore, the specific absorption rate (SAR), which is an index of the biological effect of the electromagnetic field, electric field E from the transformer of Model 2 and transmitting efficiency when the human body was included were estimated by an electromagnetic simulator. SAR and E were found to be less than the limit value prescribed by International Commission on Non-Ionizing Radiation Protection (ICNIRP) standards when the transmitting power was 15W. Difference of the transmitting efficiency between with and without a human body was less than 10.4%. This confirmed that our transformer can transmit energy to the receiving coil without adverse biological effects.
No takes yet. Share an insight, caveat, or question.
Shiba et al. (2013) studied this question.