Highlights the risk of tether failure during Micra leadless pacemaker recapture in patients with small right ventricles and demonstrates the feasibility of percutaneous recovery.
Answers are grounded in this paper and cite their sources.
Apical Micra placement in small patients may complicate recapture and risk tether damage; leaves open optimal implantation strategies.
Key Teaching Points•Device recapture during leadless pacemaker implant is frequently performed should the parameters be unsatisfactory. Operators need to cautiously align the delivery system with the deployed device before locking the tether.•In patients with small body habitus and right ventricular size, apical implantation of Micra pacemaker may increase the challenge of subsequent recaptures with the potential for damage to the tether owing to repeated abrasions.•Percutaneous recovery of an embolized Micra leadless pacemaker can be safely performed even with an unfavorable device orientation. •Device recapture during leadless pacemaker implant is frequently performed should the parameters be unsatisfactory. Operators need to cautiously align the delivery system with the deployed device before locking the tether.•In patients with small body habitus and right ventricular size, apical implantation of Micra pacemaker may increase the challenge of subsequent recaptures with the potential for damage to the tether owing to repeated abrasions.•Percutaneous recovery of an embolized Micra leadless pacemaker can be safely performed even with an unfavorable device orientation.