An imaging Raman system based on a 2-D detector and capable of collecting simultaneously 1024 spatially resolved spectra has been constructed. Hardware and software are described which allow real time analysis of the spectral data, leading to a large reduction in the data storage requirements. The analysis yields a 1-D profile (1024 points) across the sample of chemical or physical properties that are distinguishable by their Raman spectral features. Sample translations, along with repeated collection and analysis cycles, allow 2-D maps of chemical or physical properties to be generated rapidly. The spatial resolution and spectral precision are 5microm and 0.16 cm(-1) respectively. Applications to the analysis of phase transformed zones in microengineered zirconias and to measurement of an in situ temperature profile of a single carbon fiber are presented. In a typical application, 66,560 Raman spectra from an 8- x 6-mm area on a partially stabilized zirconia sample were collected and analyzed in 5.4 h to produce a 2-D map of the fraction of tetragonal phase transformed to monoclinic phase during crack propagation.
No takes yet. Share an insight, caveat, or question.
Veirs et al. (1990) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: