This is the authors' abstract. We don't add key points for this paper.
OBJECTIVES: In recent years, video-assisted thoracoscopic surgery (VATS) has increasingly become the preferred technique for thoracic surgery. However, the inherent characteristics of the lungs as large, soft, slippery, and delicate creates difficulties for pulmonary surgery. In this article, we outline the development and assessment of a balloon-based organ retractor for VATS via collaboration between medical and engineering personnel. METHODS: A dry lab trial and accompanying questionnaire assessment were performed by a group of thoracic surgeons. Objective pressure measurements were obtained, and animal experiment on pigs was performed. RESULTS: of the developed retractor was clearly lower than that for the Cherry Dissector. The questionnaire completed by the surgeons following the dry lab and animal experiments showed that most of the surgeons (7 surgeons out of 9) were satisfied with the quality of the balloon-based retractor based on a score of 3.13 ± 0.28 (mean ± standard deviation) out of 4.0. During the animal experiment, the balloon-based retractor provided stable and clear viewing with minimal need for adjustment. CONCLUSION: This balloon-based retractor could contribute to increased safety and less-invasive VATS.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Taniguchi et al. (2018) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: