Although the calibration of the flush-mounted hot film device was affected by changes in the probe elevation temperature of as little as 1 °C, it was possible to use the probe over a considerable range of elevation temperature by appropriate calibration. The behavior of the calibration coefficients with changing elevation temperature has been documented. The general probe behavior presented by Bellhouse and Shultz was substantiated and extended to a wider range of operating conditions. The experience with the mounting system designed for the probe has suggested how to obtain an improved design to maintain the probe calibration. Sensitive measurements of the wall shear stress in developing turbulent pipe flow have been taken with the probe calibrated for flow temperature changes. These appear to be consistent with trends observed by others in developing duct flow.