Does a Jarvik-7 type pneumatic artificial heart support survival and hemodynamic requirements in a calf model?
Does a Jarvik-7 type pneumatic artificial heart support survival and hemodynamic requirements in a calf model?
A Jarvik-7 type artificial heart designed for humans can sustain a calf for 221 days, demonstrating its hemodynamic capability despite eventual complications like valve pannus formation.
Supports long-term survival in growing calf model; leaves open human translation due to pannus and bleeding risks.
A Jarvik-7 type of pneumatic artificial heart, which was specifically designed to fit the anatomy and hemodynamic requirements of human patients, was implanted in a calf in an experiment to test the hemodynamic performance of the artificial heart. The experiment lasted for 221 days, longer than any animal had ever lived without its natural heart, despite the fact that the calf increased its body weight to 171 kg. The calf showed typical signs of low cardiac output before its death, but the direct cause of death was intestinal bleeding. At autopsy, it was discovered that the low cardiac output was due to severe pannus around the left inflow valve, as diagnosed earlier by changes in the pneumatic pressure wave form.
No takes yet. Share an insight, caveat, or question.
Mochizuki et al. (1981) studied this question.